Showing of 1 until 15 from 137 result(s)
Search for: Endothelium, corneal; Mitomycin; Microscopy, electron, transmission; Microscopy, electron, scanning; Aqueous humor; Vitreous body; Rabbits
Abstract
Objetivo: Este estudo prospectivo compara as imagens de microscopia confocal do epitélio corneano do coelho e do homem, obtidas através de 2 objetivas com aberturas numéricas (AN) diferentes. Métodos: Dez olhos de coelhos foram enucleados e fixados através de um suporte pneumático para garantir o melhor desempenho de cada objetiva. Cinco pacientes normais foram selecionados após consentimento. Os olhos de coelhos e dos pacientes foram previamente examinados na lâmpada de fenda. O exame de microscopia confocal (Tomey, Erlangen-Tennenlohe, Alemanha) foi realizado com as objetivas Achroplan 40x/AN = 0,75 e 63x/AN = 0,9 (Zeiss, Oberkochen, Alemanha). Imagens selecionadas do epitélio corneano foram avaliadas qualitativamente com relação ao tamanho, forma e refletividade das células. Resultados: As células no epitélio superficial dos coelhos e dos pacientes, previamente à descamação, tiveram uma refletividade maior que as células adjacentes. Este aspecto foi claramente observado somente com a objetiva 63x/AN = 0,9. As camadas basal e intermediária do epitélio em coelhos foram visualizadas somente através desta objetiva. Estas camadas nos pacientes tornaram-se mais nítidas com a objetiva de abertura numérica maior (63x/AN = 0,9). Conclusão: Uma objetiva de abertura numérica elevada produz melhor resolução dos cortes ópticos, facilitando a análise das camadas do epitélio no coelho e no homem.
Keywords: Corneal epithelium; Confocal microscopy; Epitélio corneano; Microscopia confocal
Abstract
OBJETIVO: Relatar as alterações no plexo nervoso corneano subbasal em pacientes com ceratite infecciosa de origem bacteriana utilizando a microscopia confocal in vivo.
MÉTODOS: Treze olhos de 13 pacientes com ceratite bacteriana unilateral e 12 indivíduos saudáveis como grupo controle foram incluídos prospectivamente no estudo. A microscopia confocal in vivo foi realizada em todos os pacientes em 2 momentos: na fase aguda da ceratite infecciosa e após 28 ± 0,6 meses da resolução da infecção.
RESULTADOS: A densidade dos nervos no plexo subbasal foi de 5,15 ± 1,03 mm/mm2 na fase aguda da
ceratite infecciosa (comparada com o grupo controle: 19,02 ± 1,78 mm/mm2, p<0,05). Apesar de significativa regeneração dos nervos corneanos ao longo de um intervalo de 28 meses após a resolução da infecção, a densidade dos nervos se manteve significativamente reduzida (9,73 ± 0,93 mm/mm2) quando comparada com o grupo controle (19,02 ± 1,78 mm/mm2, p<0,05). Além disso, as imagens obtidas com a microscopia confocal mostraram áreas de hiperreflectividade referente ao tecido corneano cicatricial com ramos de nervos, afinados e tortuosos, se regenerando nessas áreas.
CONCLUSÕES: Foi observado regeneração parcial dos nervos do plexo corneano subbasal durante os primeiro 28 meses após a resolução da fase aguda da ceratite infecciosa. Além disso, os nervos corneanos regenerados se mantiveram morfologicamente alterados quando comparados ao grupo controle. Esses resultados podem ser relevantes para o acompanhamento clínico e planejamento cirúrgico desses pacientes.
Keywords: Córnea/inervação; Nervo oftálmico; Infecções oculares virais; Ceratite herpética; Microscopia confocal
Abstract
OBJETIVOS: Identificar vasos linfáticos em espécimes orbitários de cadáveres humanos através de microscopia óptica e análise imunohistoquímica.
MÉTODOS: Um estudo postmortem incluiu dez espécimes orbitários provenientes de dez cadáveres humanos. Todos os espécimes orbitários foram obtidos até 12 horas após a morte com uma técnica cirúrgica de exenteração orbitária e dissecados em glândula lacrimal, nervo óptico, gordura órbitária e músculos extraoculares. Para classificar como um vaso linfático, os critérios histológicos incluíram vasos endoteliais de parede única sem membrana basal bem desenvolvida, irregulares e lúmen sem hemácias, e os critérios imunohistoquímicos incluíram vasos endoteliais de parede única, com formato irregular e lúmen sem hemácias e reagentes a podoplanina D2-40.
RESULTADOS: As lâminas histológicas de glândula lacrimal, nervo óptico, tecido adiposo e músculos extraoculares reagiram positivamente a podoplanina D2-40.
CONCLUSÃO: Este estudo demonstrou vasos linfáticos na órbita humana, mais exatamente, na glândula lacrimal, no nervo óptico, na gordura orbitária e nos músculos extrínsecos extraoculares via microscopia óptica e imunohistoquímica.
Keywords: Vasos linfáticos; Órbita; Nervo óptico; Aparelho lacrimal; Músculos oculomotores; Tecido adiposo; Microscopia
Abstract
OBJETIVO: Comparar os níveis de proteína de choque térmico 70, de periostina e de irisina no humor aquoso de pacientes com pseudoexfoliação com catarata sem glaucoma e compará-los com pacientes com catarata sem pseudoexfoliação.
MÉTODOS: Trinta e um olhos de 31 pacientes com pseudoexfoliação com catarata sem glaucoma e 30 olhos de 30 indivíduos com catarata foram incluídos neste estudo. Amostras de humor aquoso foram coletadas de todos os pacientes no momento da cirurgia de catarata e obtidas através de uma paracentese límbica por meio de uma cânula de calibre 25 acoplada a uma seringa com tuberculina. Foram coletados 100 a 150 µL de humor aquoso. Os níveis de proteína de choque térmico 70, de periostina e de irisina no humor aquoso foram medidos usando o método de ensaio imunossorvente ligado a enzima.
RESULTADOS: A média da idade (p=0,221) e sexo (p=0,530) foram semelhantes entre os grupos pseudoexfoliação e controle. Os níveis médios de proteína de choque térmico 70 foram 29,22 ± 9,46 ng/mL (17,88-74,46) e 19,03 ± 7,05 ng/ mL (9,93-35,52) nos grupos pseudoexfoliação e controle, respectivamente. Os níveis de proteína de choque térmico 70 foram maiores no grupo pseudoexfoliação (p<0,0001). O nível médio de periostina foi de 6017,32 ± 1271,79 pg/mL (3787,50-10803,57) no grupo pseudoexfoliação e 4073,63 ± 1422,79 pg/mL (2110,44-7490,64) no grupo controle. O nível médio de periostina também foi maior no grupo pseudoexfoliação (p<0,0001). Os níveis médios de irisina foram 53,77 ± 10,19 ng/mL (29,46-71,16) e 39,29 ± 13,58 ng/mL (19,41-70,56) nos grupos pseudoexfoliação e controle, respectivamente. O nível médio de irisina foi maior no grupo pseudoexfoliação do que no grupo controle (p<0,0001).
CONCLUSÕES: Os níveis de proteína de choque térmico 70, de periostina e de irisina aumentam no humor aquoso de pacientes com pseudoexfoliação sem glaucoma.
Keywords: Humor aquoso; Irisina; Proteínas de choque térmico HSP70; Periostina; Síndrome de pseudoexfoliação
Abstract
Objetivo: O sistema renina-angiotensina está envolvido na patogênese das condições isquêmicas retinianas e no glaucoma. Nosso objetivo foi avaliar a atividade da renina, enzima conversora de angiotensina 1 e 2 no humor aquoso, e amostras de sangue de pacientes com e sem glaucoma primário de ângulo aberto.
Métodos: Foram analisadas amostras de 56 participantes submetidos à cirurgia ocular. Os pacientes foram divididos em dois grupos: pacientes com catarata apenas (n=28), e pacientes com catarata e glaucoma primário de ângulo aberto (n=28). Amostras de sangue venoso (2ml) e humor aquoso (150 µl, via paracentese) foram coletadas durante a facoemulsificação (apenas catarata) ou cirurgia de glaucoma (catarata e glaucoma primário de ângulo aberto). As atividades sérica do humor aquoso de renina, enzima conversora de angiotensina 1 e enzima conversora de angiotensina 2 de todos os pacientes foram avaliadas por ensaios fluorimétricos, e os resultados foram analisados por regressão multivariada.
Resultados: Tanto a atividade da renina no humor aquoso quanto à razão humor aquoso/soro da atividade da renina foram significativamente menores nos pacientes com catarata e glaucoma primário de ângulo aberto do que em pacientes com catarata apenas [(média ± DP): 0,018 ± 0,006 ng/ml/h vs 0,045 ± 0,009 ng/ml/h; p<0,001 e 0,05 ± 0,02 vs 0,13 ± 0,05; p=0,025]. Análises multivariadas mostraram uma releção significativa entre menor atividade de renina no humor aquoso e glaucoma primário de ângulo aberto [coeficiente (±erro padrão): -0,029 ± 0,013; p=0,026].
Conclusões: Como a maioria dos pacientes com glaucoma primário de ângulo aberto usavam o colírio de timolol, estudos futuros envolvendo um maior número de pacientes e retirada prévia do tratamento são necessários para se discriminar o envolvimento do uso de betabloqueadores na atividade da renina no humor aquoso.
Keywords: Glaucoma de ângulo aberto; Catarata; Humor aquoso; Sistema renina-angiotensina
Abstract
PURPOSE: To characterize the extracellular vesicle protein cargo in the aqueous humor and plasma of patients with ocular toxoplasmosis.
METHODS: Aqueous humor and plasma were collected from six patients with active ocular toxoplasmosis and six patients with cataract. Extracellular vesicles were isolated, and western blotting and mass spectrometry were performed for protein analysis.
RESULTS: All plasma samples from patients with ocular toxoplasmosis and cataract were positive for the tetraspanins CD63 and TSG101. However, the aqueous humor from patients with ocular toxoplasmosis was positive only for CD63. Sixty-seven new unreported proteins were identified in the aqueous humor and plasma of patients with the ocular toxoplasmosis and cataract. Of the 67 proteins, 10 and 7 were found only in the cataract and ocular toxoplasmosis groups, respectively. In general, these proteins were involved in immune system activation and retina homeostasis and were related to infections and retina-associated diseases. Conclusion: The distinct protein signatures between ocular toxoplasmosis and cataract may be helpful in the differential diagnosis of ocular toxoplasmosis. However, more studies are needed to better understand the role of these proteins in the pathogenesis of ocular toxoplasmosis.
Keywords: Extracellular vesicles; Proteomics; Toxoplasma gondii; Ocular toxoplasmosis, Aqueous humor; Plasma; Liquid biopsy
Abstract
MÉTODOS: Córneas humanas de treinamento disponibilizadas foram randomizadas em quatro grupos: Pachy-100 (profundidade de incisão = espessura corneana central - margem de segurança de 100 µm), Pachy-50 (margem de segurança de 50 µm), Pachy-0 (sem margem de segurança) e Pachy+50 (profundidade de incisão = espessura corneana central + 50 µm). Todas as lamelas foram dissecadas através um método padronizado e já publicado (Pachy-DSEK). As espessuras das lamelas (centro, zona de 3,0mm e zona de 6,0mm) foram medidas com tomografia de coerência óptica. A razão de espessura centro-periferia foi calculada aos 3,0 e 6,0 mm de diâmetro.
RESULTADOS: Perfuração endotelial ocorreu apenas no grupo Pachy+50 (n=3, 30%). A espessura central da lamela nos grupos Pachy-100, Pachy-50, Pachy-0 e Pachy+50 foi de 185 ± 42 µm, 122 ± 29 µm, 114 ± 29 µm, e 58 ± 31 µm, respectivamente (p<0,001). As razões C/P aos 3,0 e 6,0 mm foram de 0,97 ± 0,06 e 0,92 ± 0,14, respectivamente. Os parâmetros de características do doador não se correlacionaram com os resultados de espessura de lamela. A profundidade planejada de incisão se correlacionou com a maioria dos parâmetros de espessura de lamela (p<0,001). A espessura de lamela se correlacionou negativamente com a profundidade planejada da incisão (p<0.001, r=-0,580). O melhor resultado foi observado no grupo Pachy-0, em que 75% das lamelas mediram abaixo de 130 µm e não houve perfuração endotelial.
CONCLUSÃO: Através de um método padronizado de dissecção, a maioria das lamelas endoteliais apresentou uma configuração planar. O planejamento de profundidade de incisão igual à espessura corneana central resultou em alta porcentagem de lamelas ultrafinas sem ocorrência de perfuração.
Keywords: Transplante de córnea; Ceratoplastia lamelar; Endotélio corneano; Dissecção; Tomografia de coerência óptica
Abstract
PURPOSE: To evaluate the reliability and comparability of a Scheimpflug-based tomographer relative to a Placido-based topographer and specular microscopy in healthy eyes.
METHODS: This cross-sectional study included 40 patients (80 eyes). Each eye underwent randomized imaging with a Scheimpflug-based tomographer, a Placido-based topographer, and Tomey EM-4000 specular microscopy. Three acquisitions per device were obtained. For interdevice comparisons, the best-quality scan per eye/device was selected, whereas all three scans were used for intradevice repeatability analyses. Unreliable scans were repeated (up to five attempts) and excluded if acceptable quality was not achieved, resulting in variable denominators. Between-device comparisons were performed using generalized estimating equations
with participant-level clustering and robust standard errors and were supplemented by Bland–Altman analysis.
RESULTS: The effective sample size varied by parameter (independent summaries: 59–67 eyes; paired comparisons: 48–51 eyes). In paired-eye analyses, the Scheimpflug-based tomographer measured slightly higher keratometry values than the Placido-based topographer (K1: 43.95 vs. 43.78 D, p=0.003; K2: 44.91 vs 44.73 D, p=0.002), more negative Q-values (p=0.001), smaller photopic pupil diameter (p<0.001), and shorter kappa distance (p<0.001). Mean absolute differences were 0.32 D for K1 and 0.30 D for K2, with high dispersion for angular metrics (kappa angle coefficient of variation: 195%).
CONCLUSIONS: The Scheimpflug-based tomographer provides reproducible corneal measurements in healthy eyes. However, systematic differences relative to the Placido-based topographer—particularly for keratometry, asphericity, and pupil and kappa metrics—suggest limited interchangeability. Consistent device use is recommended when these parameters inform clinical decision-making.
Keywords: Scheimpflug tomography; Placido topography; Specular microscopy; keratometry; Corneal imaging; Refractive surgical procedures; Lenses, intraocular
Abstract
PURPOSE: To evaluate the effect of single-session transscleral diode laser cyclophotocoagulation on intraocular pressure in refractory glaucoma and to determine structural changes using ultrasound biomicroscopy.
METHODS: Forty-three eyes were evaluated. Intraocular pressures at baseline and at the first, third, and sixth months after transscleral diode laser cyclophotocoagulation were compared. Ciliary body thickness, ciliary muscle thickness, ciliary process thickness, iris root thickness, and scleral thickness were assessed at baseline and at the third and sixth months post-treatment.
RESULTS: Reductions in intraocular pressure were significant between baseline and the first month (p=0.018), third month (p<0.001), and sixth month (p<0.001) as well as between the first and third months (p=0.034) and the first and sixth months (p=0.036). Compared with baseline, intraocular pressure reduction rates at the first, third, and sixth months were 34.6%, 56.5%, and 55.3%, respectively, while success rates were 30.2%, 62.8%, and 55.8%, respectively. Decreases in ciliary body thickness, ciliary muscle thickness, and ciliary process thickness were significant between baseline and the third month (p<0.05) and between baseline and the sixth month (p<0.05), whereas changes between the third and sixth months were not significant (p>0.05). Iris root and scleral thicknesses did not change after treatment (p>0.05). At the third and sixth months, significant positive correlations were observed between changes in intraocular pressure and changes in ciliary body thickness and ciliary process thickness (p<0.05).
CONCLUSIONS: To the best of our knowledge, this is one of the few studies comprehensively investigating structural changes after transscleral diode laser cyclophotocoagulation using ultrasound biomicroscopy. Moreover, the relationships between intraocular pressure changes and variations in the ciliary body, ciliary muscle, ciliary process, iris root, and scleral thicknesses were examined in detail. Single-session treatment did not affect iris root or scleral thickness but significantly reduced ciliary body, ciliary muscle, and ciliary process thicknesses. Greater reductions in ciliary body and ciliary process thickness may contribute to more pronounced intraocular pressure reduction.
Keywords: Intraocular pressure; Laser coagulation/methods; Lasers, semiconductor; Microscopy, acoustic; Glaucoma; Ciliary body
Abstract
Objetivo: O efeito regulador do microRNA em doenças tem sido confirmado, e este artigo tentou avaliar a expressão do microRNA-210-3p na catarata relacionada à idade e avaliar o efeito da expressão anormal do miR-210-3p em células SAR01/04 induzidas por H2O2.
Métodos: O método de transcrição reversa seguida de reação em cadeia da polimerase (RT-PCR) quantitativa foi realizado para avaliar os níveis de miR-210-3p em amostras de humor aquoso. Análise de características operacionais do receptor foi feita para avaliar a capacidade de discriminação do miR-210-3p entre pacientes com catarata relacionada à idade e pessoas saudáveis. A análise de correlação de Pearson identificou a correlação do miR-210-3p e índices de estresse oxidativo, como superóxido dismutase, glutationa peroxidase, malonaldeído. O ensaio de contagem de células kit-8 (cck-8) e o ensaio no sistema Transwell foram utilizados para estimar a função biológica do formato de células de catarata relacionada com a idade induzida por H2O2. Os níveis de índices de estresse oxidativo como superóxido dismutase, glutationa peroxidase e malonaldeído foram detectados para avaliar o grau de dano do estresse oxidativo em formato de células de catarata relacionada à idade. A relação entre miR-210-3p e seu gene alvo foi verificada por análise do gene repórter luciferase.
Resultados: A expressão miR-210-3p foi elevada no humor aquoso de pacientes com catarata relacionada à idade. A expressão miR-210-3p altamente expressiva mostrou alto valor diagnóstico para catarata relacionada à idade e foi significativamente associado ao nível de marcadores de estresse oxidativo em pacientes com catarata relacionada à idade. A inibição de miR-210-3p pode reverter a estimulação do estresse oxidativo e os efeitos adversos da função celular induzida por H2O2.
Conclusões: Esses dados sugeriram que a expressão miR-210-3p poderia promover a viabilidade celular, migração celular e estresse oxidativo ao direcionar genes ATG 7 relacionados à autofagia em modelo in vitro de células de catarata relacionadas à idade.
Keywords: Catarata; Fatores etários; Humor aquoso; MiR-210- 3p; Estresse oxidativo; Proteína 7 relacionada à autofagia
Abstract
PURPOSE: This study aimed to evaluate the efficacy and clinical outcomes of a one-way fluid-air exchange procedure for the treatment of postvitrectomy diabetic vitreous hemorrhage in patients with proliferative diabetic retinopathy.
METHODS: This retrospective study included 233 patients with proliferative diabetic retinopathy, who underwent vitrectomy. A one-way fluid-air exchange procedure was performed in 24 eyes of 24 (10.30%) patients with persistent vitreous cavity rebleeding after the operation. Preprocedural and postprocedural best-corrected visual acuity values were achieved. Complications occurring during and after the procedure were analyzed.
RESULTS: Significant visual improvement was observed 1 month after the one-way fluid-air exchange procedure (2.62 ± 0.60 LogMAR at baseline vs. 0.85 ± 0.94 LogMAR at postprocedure, p<0.0001). Moreover, 19 (79.17%) eyes needed the procedure once, and 5 (20.83%) eyed had the procedure more than twice. In 3 (12.50%) eyes, reoperation was eventually required because of persistent rebleeding despite several fluid-air exchanges. No complication was observed during the follow-up.
CONCLUSIONS: The one-way fluid-air exchange procedure can be an excellent alternative to re-vitrectomy for patients with proliferative diabetic retinopathy suffering from postvitrectomy diabetic vitreous hemorrhage by removing the hemorrhagic contents directly and achieving fast recovery of visual function without apparent complications.
Keywords: Diabetic retinopathy; Vitrectomy; Vitreous body; Vitreous hemorrhage; Hemostatic techniques
Abstract
Trata-se de uma paciente do sexo feminino, de 45 anos, portadora de ceratocone, submetida a uma ceratoplastia endotelial com membrana Descemet após apresentar um quadro de perda de visão severa devido a uma hidrópsia corneana aguda no olho esquerdo. Inicialmente, a acuidade visual corrigida da paciente era de 20/80 no olho direito e de movimento de mãos no olho esquerdo. Após exame de biomicroscopia que detectou uma extensa rotura da membrana de Descemet e edema estromal, optamos por tratar esse caso com o ceratoplastia endotelial com membrana Descemet. Doze meses após o procedimento cirúrgico, percebeu-se uma melhora do edema corneano, não havia sinais de rejeição do botão óptico e a acuidade visual corrigida da paciente era de 20/50 no olho afetado.
Keywords: Ceratocone/cirurgia; Ceratoplastia endotelial com remoção da lâmina limitante posterior Microscopia com lâmpada de fenda
Abstract
Um homem de 60 anos, diagnosticado com mieloma múltiplo recidivante refratário, apresentou vermelhidão, dor, sensação de corpo estranho e visão turva em ambos os olhos, aumentando gradualmente após sua terceira infusão de belantamabe mafodotina. À biomicroscopia, foram observadas alterações epiteliais bilaterais semelhantes a microcistos e depósitos epiteliais semelhantes a cristais. A microscopia confocal in vivo revelou depósitos hiper-refletivos intraepiteliais e subepiteliais na córnea. Devido à toxicidade corneana, a terapia com belantamabe mafodotina foi interrompida por sete semanas e a toxicidade foi gradualmente resolvida. Nosso objetivo é demonstrar os achados à biomicroscopia confocal in vivo e à lâmpada de fenda da toxicidade corneana relacionada ao belantamabe mafodotina.
Keywords: Mieloma múltiplo; Belantamabe mafodotina; Microscopia confocal; Córnea; Efeitos colaterais e reações adversas relacionados a medicamentos
Abstract
Neonato de 7 semanas, do sexo masculino, nascido de parto cesárea, apresentou sorologia positiva para dengue com hemorragias retinianas e pré-retinianas, opacidades vítreas e manchas algodonosas. O paciente e sua mãe haviam apresentado sorologias positivas para Non Structural Protein 1 através de ELISA. Achados na retina e no vítreo melhoraram em um período de dezesseis semanas. O exame de tomografia de coerência óptica de domínio espectral demonstrou arquitetura macular preservada. Neste relato de caso, sugerimos que alterações na retina e no vítreo podem ser os achados oculares aparentes em neonatos com infecção vertical por dengue, e que estes podem se resolver sem maiores sequelas estruturais.
Keywords: Dengue; Complicações infecciosas na gravidez; Transmissão vertical de doenças infecciosas; Manifestações oculares; Hemorragia retiniana; Hemorragia vítrea; Humanos; Doenças do recém-nascido; Relatos de casos
Abstract
A ocorrência de ectasia corneana após ceratectomia fotorrefrativa é uma complicação rara, porém grave, em cirurgia refrativa. Os possíveis fatores de risco não são bem avaliados, mas a opinião atual é que a falha na detecção de ceratocone pré-operatório possa ser o principal motivo. Neste relato, descrevemos um caso de ectasia corneana após ceratectomia fotorrefrativa em paciente apresentando padrão tomográfico suspeito no pré-operatório, mas sem alterações degenerativas associadas a ceratocone patológico, conforme revelado por microscopia confocal in vivo da córnea. Além disso, revisamos, na literatura, relatos de casos elegíveis de ectasia pós-ceratectomia fotorrefrativa para encontrar características semelhantes.
Keywords: Córnea/patologia; Dilatação patológica; Ceratoconus; Ceratectomia fotorrefrativa; Ceratomileuse assistida por excimer laser in situ; Microscopia confocal; Humanos; Relato de casos